Typical applications
The default application is to heat the CompreCell 6/!2 or CompreCell 12-DP to allow for temperature dependent measurements:
Determination of the temperature-dependent conductivity of solid electrolytes
Cycling solid-state batteries at different temperatures
Stripping-plating experiments at different temperatures
Selected views and impressions
Detailed Description
The H add-on is compatible with the CompreCell 6/12 and the CompreCell 12-DP. It is not compatible with the CompreCell Pouch. It is designed to enable the electrochemical testing of solid electrolytes or all-solid-state batteries while precisely regulating the applied temperature of the sample between room temperature and +250 °C.
The heater can be adjusted to ensure best thermal contact to the cell. Moreover, the heater has a built-in insulation to lower its outer temperature. It is also connected to protective earth to provide shielding of the cell from electromagnetic noise. In addition, the GND connector of the measuring device can be connected to the heater as well.
The CompreDrive H add-on is equipped with a serial connection interface which enables the communication between any PC and the temperature controller. Using this interface, automated temperature-dependent measuring procedures can be established.
The CompreDrive HC add-on is fully implemented in CompreDriveControl enabling automated temperature adjustment during measuring procedures.
Technical Details
- Mains voltage:
UAC (rms) = 115 to 230 V, 50-60 Hz
- Mains fuse:
5 A
- Heater voltage:
UDC = 48 V
- Temperature range during operation (measuring cells):
Toperation = room temperature to +250 °C
- Temperature range during operation (environment):
Tenv. operation = +10 °C to +40 °C
- Temperature range during storage:
Tstorage = +10 °C to +40 °C
- Relative humidity (RH) for working and storage area:
(non-condensing) 0 to 80% RH
- Atmosphere during storage:
non-corrosive
- Rated dimensions of the temperature controller unit:
33 cm x 19 cm x 16 cm (L x W x H)
- Total weight:
5.6 kg
Key Data
- Art. No.:
840107
- Essential features at a glance:
- Temperature range during operation (measuring cells): Toperation = RT to +250 °C
- Mains voltage: UAC (rms) = 115 to 230 V, 50-60 Hz
- Embedded serial communication to temperature controller
- Compatible with CompreDriveControl software