Typical applications
Understanding the performance of battery cells during cycling is a critical factor. If mechanical characteristics are not taken into account, battery research can end up with irreproducible results and, even worse, with issues when trying to transfer the obtained results to the module and pack level. Typical applications with the isostatic pouch cell testing fixture CompreStat are:
Cycling of pouch cells at different isostatic pressures (up to 3 pouch cells at once)
Pressurized cycling at various temperatures (up to 3 pouch cells at once)
Improving reproducibility by well-defined pressure and temperature conditions
Enable unprecedented performance of all solid state cells
Selected views and impressions
Detailed Description
The CompreStat is a pressurized cell fixture that enables electrochemical tests of batteries in a pouch cell format at different isostatic pressure and temperature conditions. Up to three cells of various formats and tab configurations can be connected independently.
One pressure sensor in the container as well as one temperature sensor per cell are integrated. The included sensor electronics and software make it a turn-key system with full control of all relevant data. It is monitored and stored via a connected PC. You can use all data handling and plotting features that our CompreDriveControl software offers. The sample temperature can be controlled by placing the CompreStat inside a temperature chamber.
Pressure is applied by gaseous media. Simply connect to Argon bottles for example, using the available filling set. No oil, no spills, no cleaning.
Safety first: A pressure relief valve is included. The product is safe by design, in case of failure of the pressure relief valve, the lid will leak in a defined way. Runaway of 3 cells is possible without bursting of the container.
Technical Details
- Maximum sample outer dimensions excl. tabs (for same side pouch cells):
150 mm x 110 mm x 10 mm
- Maximum sample outer dimensions excl. tabs(for opposite side pouch cells):
135 mm x 110 mm x 10 mm
- Pressure range during operation (CompreStat):
poperation = 0 to 10 MPa (rel)
- Filling medium:
Gas
- Built-in pressure sensor – Measuring range:
0 . . . 12 MPa
- Pressure resolution:
0.01 MPa
- Reference pressure:
Ambient pressure
- Characteristic deviation according to IEC 61298-2:
+/- 0.05 MPa
- Temperature range during operation (electronics + accessories):
Toperation = +10 °C to +40 °C
- Maximum measurement potential:
40 V ac/dc
- Maximum measurement current (using supplied contact solution):
10 A
- Rated dimensions :
CompreStat: 309 mm x 140 mm x 135 mm
CompreStat Data Sampler: 140 mm x 106 mm x 60 mm
- Typical weight:
CompreStat: 6,0 kg
- Connection for pressurized gas cylinders:
W30x2” (DIN 477-5)
Key Data
- Art. No.:
840130
- Essential features at a glance:
- High quality setup for testing pouch cells under isostatic pressure conditions
- Pressure and temperature monitoring
- Same-side and opposite-side tab connector solutions
- Large pressure range up to 10 MPa with 0.01 MPa resolution
- Large temperature range when being operated in climated chamber: -23 °C … + 80 °C
- Universal compatibility with all potentiostats/galvanostats/battery cyclers