Typical applications
This test cell can be used for experiments with both a 2-electrode and a 3-electrode configuration. Typical applications are:
- Determination of the conductivity of liquid electrolytes
- Determination of the electrochemical stability window of electrolytes
- Cyclic voltammetry studies of sensitive substances
Selected views and impressions
Detailed Description
The test cell TSC 1600 Closed GC consists of several high-quality building blocks: The gold-plated base unit (= “thermo block”) comes with an integrated Pt100 temperature sensor that monitors the temperature very close to the sample, as well as a fixed platinum crucible that serves as sample container and counter electrode. To seal the test cell, a housing made of polyether ether ketone (PEEK) with two O-ring grooves is used which connects the base unit and the cap.
In case of the TSC 1600 Closed GC, the PEEK cap has an exchangeable electrode tip made of a glassy carbon rod press-fitted in a PEEK sleeve. The mirror-like polished surface of the glassy carbon rod is exposed to the sample electrolyte. Other electrode tip materials are available on demand.
Furthermore, the cap has two inlets that are sealed by plugs by default, but can also be used for inserting capillaries or (pseudo-) reference electrodes made of metal wires.
Technical Details
- Capacity:
max. 1.0 ml, min. 0.9 ml
- Test cell materials:
platinum (sample container and electrode), glassy carbon (electrode tip), gold-plated copper with nickel interlayer (base unit), PEEK (housing, cap, and electrode tip), EPDM, FKM, or FFKM (O-rings)
- Sensor:
Integrated Pt100 temperature sensor
- Electrode tip (being part of the cap; exchangeable):
3 mm diameter glassy carbon rod press-fitted in PEEK, front-end mirror-like polished, contacted from the backside by a gold-plated contact spring
- Temperature range (operation):
-40 °C to +100 °C depending on sample amount and dew point
- Temperature range (storage):
+10 °C to +30 °C
- Required accessory to make electrical connection:
Microcell HC Setup, Microcell Passive, or customized cable set
Key Data
- Art. No.:
840217
- Material in contact with sample:
platinum, PEEK, glassy carbon, O-ring material (default material = EPDM, FFKM)
- Working electrode material:
glassy carbon
- Counter electrode material:
platinum
- Sample volume:
min. 0.9 ml