Typical applications
The CompreFrame can be used to operate the test cells CompreCell 12, CompreCell 12-DP and CompreCell Pouch. Therefore, a broad range of different applications is accessible:
Cycling of solid-state batteries at precisely adjusted stack pressure and under active temperature control
Cycling of pouch cell formats with lithium-ion battery or solid-state battery chemistry at precisely adjusted stack pressure and under temperature control
Pressure- and temperature-dependent investigation of a solid-state electrolyte’s conductivity
Stripping-plating experiments under actively controlled stack pressure
Dilatation monitoring during cycling of pouch cells
Selected views and impressions
Detailed Description
The CompreFrame is a cell stand for all types of CompreCells (CompreCell 6/12, CompreCell 12 DP and CompreCell Pouch) enabling very precise manual pressure application to solid electrolyte powders, pellets, all-solid-state batteries (ASSB), and pouch cells.
A display unit always shows the current force, making stand-alone operation possible. The force and sample temperature can also be recorded and monitored using the included CompreDriveControl software.
The CompreFrame offers automated temperature control, if used with the optional H- or HC Add-On. Different pressure sensors are available for best accuracy. If other spring assemblies should be required, please contact us to discuss customization options.
Technical Details
- Maximum applicable force onto measuring cells:
Fmax = 30 kN
Maximum pressure depends on sample area:
pmax = 265 MPa with 12 mm pistons and 1 GPa with 6 mm pistons in CompreCell 12/6 or CompreCell 12DP/6DP*
pmax = f(pouch cell area) in CompreCell Pouch
- Minimum applicable force onto measuring cells:
Fmin = 0.3 kN
Minimum pressure depends on sample area:
pmin = 2.7 MPa with 12 mm pistons and 10 MPa with 6 mm pistons in CompreCell 12/6 or CompreCell 12DP/6DP*
pmax = f(pouch cell area) in CompreCell Pouch
- Torque needed for pressure adjustment:
for 20 kN / 30 kN: 35 Nm / 50 Nm
- Force needed for pressure adjustment:
for 20 kN / 30 kN at lever: 90 N / 130 N
- Sensor accuracy:
0.1% of full scale in all cases
- Available force sensors:
100 kN, 10 kN, 2 kN maximum force (0.1% linearity)
- Accessible temperature range during operation (measuring cell):
CompreCell 12/6 or CompreCell 12DP/6DP:
Toperation/cell = -40 °C to +250 °C** with HC-Addon and RT to +250 °C with H-Addon
CompreCell Pouch:
Toperation/cell = 0 °C to +100 °C with HC-Addon
- Temperature range (environment) during operation (CompreFrame):
- Temperature range (environment) during operation (accessories):
- Temperature range for storage:
- Relative humidity range during operation and storage:
0 to 80% (non-condensing)
- Atmosphere during storage:
non-corrosive
- Rated dimensions (CompreFrame):
19.2 cm x 19.0 cm x 51.4 cm (L x W x H)
- Rated dimensions (Controller):
14.0 cm x 10.6 cm x 6.0 cm (L x W x H)
- Total weight:
approx. 16.5 kg
- Control software:
CompreDriveControl / CompreFrameControl
*The max. force might be limited by the selected force sensor and the chosen test cell.
**The accessible range is limited by the coupled thermostat.
Key Data
- Art. No.:
840115
- Essential features at a glance:
- High quality setup for solid-state research
- Compatibility with all test cells with prefix CompreCell
- Airtight measuring cells allow for the handling of sensitive samples outside of a glovebox
- Large range of applicable force up to 30 kN
- Large range of sample temperature from -40 °C up to 250 °C
- Universal compatibility with all potentiostats/galvanostats/battery cyclers
- Various capabilities for inter-device communication in hardware and software for almost completely automated electrochemical routines with most potentiostat/galvanostat manufacturers
- Stand-alone operation possible
Related application notes
- News and insights
New Publication: Si Nanowire All-Solid-State Battery
- Application notes
Lithium Metal Solid-State Pouch Cell: Pressure Dependence of Rate Performance
- Application notes